1.
It only takes 6-7 minutes for the shortest sterilization cycle which can extend
service life of the instrument
2.
Three shortcut keys for sterilization programs, the sterilization temperature
and time can be set freely
3.
Alarm for unqualified water and water shortage, the autoclave will stop working
when there is a low temperature and low pressure
4.
Small dimension, light weight, easy to operate, the cassette can be also used
for instrument transfer and temporary storage
Schematic
Diagram:
The
steam generator heats to the optimum temperature when the sterilization program
started. The distilled water is pumped into the generator and converted into
steam immediately, then the steam is injected into the cassette which is loaded
with instruments to be sterilized. The polluted air in the cassette is exhausted
after the steam is injected.
Technical Parameters:
Model | BKS-2000 | BKS-5000 | BKS-6000 |
Capacity | 1.8L | 5.2L | 6.0L |
Chamber
Size(W*D*H)mm | 280*180*38 | 380*180*78 | 480*180*78 |
External
Size(W*D*H)mm | 570*415*170 | 580*460*190 |
Temp.
Range | 115℃~135℃ |
Temp.
Fluctuation | ≤±5℃ |
Ambient
Temp. | 5~40℃ |
Working
Pressure | 42
kPa ~212kPa |
Chamber
Material | 304
Stainless Steel |
Water
Tank Volume | 3.4
L |
Standard
Accessories | Instrument
loading tray*1, Waste water tank*1 Drain
hose*1, Printing paper*2 |
Power
Supply | AC
220V/110V±10%, 50Hz/60Hz |
Packing
Size(W*D*H)mm | 580*580*700 |
Net
Weight(kg) | 34 | 46 |
Gross
Weight(kg) | 49 | 61 | 61
|
Standard
Accessories:
Instrument
loading tray*1, Waste water tank*1, Drain hose*1, Printing paper*2 |