BK-D560/BK-D580/BK-D590 DOUBLE BEAM SCANNING UV/VIS SPECTROPHOTOMETER

  • FOB Price:Get Latest Price >
  • Min.Order:1 Set(s)
  • Payment Terms:L/C , T/T , PayPal , Money Gram , EXW , FCA
  • Favorite
Biobase Biodustry (Shandong) Co., Ltd

Business Type:Manufacturer

Country/Region:China

Ddu Verified

HOT Rank

8/10

Product Information

  • Model:BK-D560
  • Certification:CE
  • Net Weight:23kg
  • Dimension:740*570*440mm
  • Packaging Details:International Standard
  • Brand Name:BIOBASE
  • Trademark:BIOBASE
  • Place of Origin:Shandong,China(Mainland)
  • Warranty:1 year

Description

Features:


⊙Environmental Deuterium lamp, preventing from Ozone inhalation

⊙Real-time monitoring the lifetime of Deuterium lamp and Tungsten lamp

⊙Socket-Type Deuterium lamp and Tungsten lamp

GLP self-check function:check the wavelength accuracy and Photometry accuracy,provide test report and power-off protection


⊙Wavelength calibration, wavelength setting, change lamp source and dark current calibration automatically

⊙Si02 coating optical mirror, reduces the influence from outside 



 

Basic Functions:

1, Photometry

Test Abs., Transmittance and Energy by fixed wavelength 

2, Quantitative

Linear fit and Linear fit through zero two modes

 a.Coefficient, Standard Sample input and Standard Sample read three modes to establish standard curve    

 b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings  

 c. Can save 240 group curves, can test 240 data in each curve 

 d. Double wavelength, Triple wavelength test functions.

3, Kinetics

Used for time course scanning or reaction rate calculations△A/t, can search all data 

4, Multi-Wavelength

Can test Transmittance and Abs. with 8 different wavelengths at most

5, Scanning

User can set the scan range from 190nm to 1100nm to test the max. Abs. peak value, can do derivation,

arithmetical operationsto the graph.

6, Biology

6 methods: DNA/Protein, UV, Lowry, BCA,CBB and Biuret

Model

BK-D560

BK-D580

BK-D590

Optical System

Double Beam, Grating 1200 lines/mm

Wavelength Range

190~1100nm

Spectral Range

1.8nm

1nm

0.5,1,2,4,5nm

Wavelength Accuracy

± 0.3nm

Wavelength Repeatability

≤ 0.2nm

Photometric Accuracy

± 0.002A (0~0.5Abs), ± 0.004A (0.5~1.0Abs), ± 0.3% T (0~100% T)  

Photometric Repeatability

0.001Abs (0~0.5Abs), 0.002Abs (0.5~1.0Abs), ≤ 0.2% T (0~100% T)

Stray Light

≤ 0.05 T @ 360nm; 220nm

Stability

± 0.0008A / h @ 500nm

Baseline Flatness

± 0.0005A

Noise

± 0.0002Abs

Display

65 thousand true color 7 inch TFT LCD(480 *800)

Photometric Mode

T, A, C, E

Photometric Range

0~200% T,-0.301~3.0A

Detector

Silicon Photodiode

Light Source

Deuterium Lamp, Tungsten Lamp

Input

Membrane Keypad

Output

USB  Print and data output, Connect PC

Standard Accessories

10mm glass cuvette *4 units  10mm quartz cuvette *2 units,software

Optional Accessories

Single cell holder,General cell holder,Test tube holder,Solid sample holder,Deuterium lamp,Halogen lamp,4-place Cell holder,Printer,Water-jacketedcell holder,Reflection accessory 5°,Auto 8-cell changer,Micro cell holder,Water-jacketed cell holder,Constant-temperature system

 

Power Supply

AC110V/220±10%, 60/50Hz

Packing Size(mm)

740*570*440

Gross Weight(kg)

25